Background/purpose: Matching phasic pressure tracings between a fluid-filled catheter and high-fidelity pressure wire has received limited attention, although each part contributes half of the information to clinical decisions. We aimed to study the impact of a novel and automated method for improving the phasic calibration of a fluid-filled catheter by accounting for its oscillatory behavior.
Â
Methods/materials: Retrospective analysis of drift check tracings was performed using our algorithm that corrects for mean difference (offset), temporal delays (timing), differential sensitivity of the manifold transducer and pressure wire sensor (gain), and the oscillatory behavior of the fluid-filled catheter described by its resonant frequency and damping factor (how quickly oscillations disappear after a change in pressure).
Â
Results: Among 2886 cases, correcting for oscillations showed a large improvement in 28% and a medium improvement in 41% (decrease in root mean square error >0.5 ...