five

Preliminary results of cervical impedance measurements

收藏
DataCite Commons2020-08-29 更新2024-07-27 收录
下载链接:
https://scielo.figshare.com/articles/Preliminary_results_of_cervical_impedance_measurements/6317990/1
下载链接
链接失效反馈
官方服务:
资源简介:
AbstractIntroduction: In this work, we report the development and initial tests of a measuring system that is used to measure the cervical electrical impedance with the purpose of assessing cervical ripening. The objective is to offer a system that provides real-time information about the cervical electrical impedance and, indirectly, of the cervical ripening with the aim of preventing premature birth. Methods Probes were developed and initial tests were performed. The probes used had electrodes that are far apart, with the assumption that in such cases deeper layers of the cervical tissue are measured. Results Preliminary measurements were performed in a single session on a non-gravid uterus, using as excitation signal a sinusoidal wave with frequency varying from 3 kHz to 100 kHz. Results indicate, as expected, that the magnitude of the impedance is inversely proportional to the frequency, indicating capacitive impedance. Conclusion Preliminary results were very encouraging. However, further measurements must be performed to correlate the impedance to the cervical ripening.