Concurrent validity of handgrip strength between the jamar and bulb dynamometers in women with rheumatoid arthitis
收藏Figshare2020-02-01 更新2026-04-28 收录
下载链接:
https://figshare.com/articles/dataset/Concurrent_validity_of_handgrip_strength_between_the_jamar_and_bulb_dynamometers_in_women_with_rheumatoid_arthitis/11869083
下载链接
链接失效反馈官方服务:
资源简介:
Abstract Introduction: Rheumatoid arthritis (RA) is a progressive disease that causes deformation and constant joint damage. Handgrip strength (HGS) has been used by several health professionals in clinical practice as a mechanism for assessing muscle strength and overall performance. Objective: To perform the concordance analysis of handgrip strength measurements using a pneumatic dynamometer (Bulb) and a hydraulic dynamometer (Jamar) in women with rheumatoid arthritis. Method: The HGS measurements by the two dynamometers followed the norms of the American Society of Hand Therapists. The concordance between measurements was performed by the Bland-Altman method. Significance level was 5%. Results: In total, 41 women (60.63 ± 8.35 years) participated in the study. Analysis showed that the measures between the two dynamometers were not concordant (bias = 9.04, p