Comparative analysis of postural control and vertical jump performance between three different measurement devices (PlosOne)

收藏
DataCite Commons2020-08-27 更新2024-07-27 收录
下载链接:
https://figshare.com/articles/Comparative_analysis_of_postural_control_and_vertical_jump_performance_between_three_different_measurement_devices_PlosOne_/7739309
下载链接
链接失效反馈
官方服务:
资源简介:
Objectives: The aim of this study was to examine the concurrent validity of the HUMAC Balance System and Balance Trainer BTG4 in comparison to a laboratory-grade force platform for postural control (PC) and vertical jump performance (VJP) assessment, and for PC to additionally explore reliability of these devices. Methods: Overall 22 participants (age = 37.8 ± 13.3 years; gender = 9 male, 13 female; height = 174.1 ± 10.5 cm; body mass = 75.3 ± 17.6 kg) were recruited to participate. Double and single leg standing balance tests with eyes open or closed and counter movement jumps (CMJ) were performed on two separate occasions. Reliability and concurrent validity for COP parameters and VJP were examined using intraclass correlation coefficients (ICC), Bland-Altman plots (BAP), standard error of measurement (SEM) and minimum detectable change (MDC). Results: COP path length test-retest reliability was predominantly good to excellent for all three devices (ICC = 0.80-0.95). SEM and MDC values were high for all plates (SEM% = 8.0-15.2; MDC% = 22.8-44.5), with the HBS MDC values higher than the KIS and BTG in three of the four trials. ICC scores for concurrent validity were good to excellent for the BTG (ICC = 0.76-0.93) and moderate to good for the HBS (0.49-0.83). Band-Altman plots revealed a systematic bias for the HBS towards higher COP path length values under all conditions and for the BTG in two out of four trials towards lower values. Validity of VJP was excellent for the BTG (ICC = 1.0) and poor for the HBS (0.34), with a systematic bias towards lower values. Conclusion: The comparative analysis of postural control and VJP revealed reliable and valid results for the BTG in comparison to a laboratory-grade force plate. The HBS showed reliable results for PC assessment with restrictions regarding its validity. Results of VJP showed that the HBS revealed deficits in the assessment of activities that require rapid, high force movements such as jumping and running. Due to the variable results of all three devices, it is recommended not to use them interchangeably.
提供机构:
figshare
创建时间:
2019-02-19
5,000+
优质数据集
54 个
任务类型
进入经典数据集
二维码
社区交流群

面向社区/商业的数据集话题

二维码
科研交流群

面向高校/科研机构的开源数据集话题

数据驱动未来

携手共赢发展

商业合作