Electrochemistry Boot Camp CV Data
收藏NIAID Data Ecosystem2026-05-02 收录
下载链接:
https://zenodo.org/record/11179881
下载链接
链接失效反馈官方服务:
资源简介:
Data were collected on 09/20/2012
The electrochemical cell consisted of:
Pt/Ir working electrode
Ag/AgCl reference electrode
Pt wire counter electrode
Scan rates of 20mV/s, 50 mV/s, 100mV/s, 150 mV/s, and 200mV/s were tested
A 10 sec quiet time was used for all scans (stir bar off after mixing for 30 sec)
The solution was composed of:
potassium ferrocyanide (4 mM) as redox probe
potassium nitrate (1 mM) as electrolyte
The testing conditions were:
Temp = 21C (room temperature)
pH=7.10
A BASi Epsilon potentiostat was used for all experiments
Details and further analysis are shown in Vanegas et al (2015)
link: https://pubs.rsc.org/en/content/articlelanding/2014/an/c4an01788a/unauth