Inaccuracy of venous point-of-care glucose measurements in critically ill patients: a cross-sectional study

收藏
DataONE2020-06-30 更新2025-04-19 收录
下载链接:
https://search.dataone.org/view/sha256:756abd4bf6ddbbfed55528e9e70440b62fdb11a224671e83aa4f811b9230c68c
下载链接
链接失效反馈
官方服务:
资源简介:
Introduction: Current guidelines and consensus recommend arterial and venous samples as equally acceptable for blood glucose assessment in point-of-care devices, but there is limited evidence to support this recommendation. We evaluated the accuracy of two devices for bedside point-of-care blood glucose measurements using arterial, fingerstick and catheter venous blood samples in ICU patients, and assessed which factors could impair their accuracy. Methods: 145 patients from a 41-bed adult mixed-ICU, in a tertiary care hospital were prospectively enrolled. Fingerstick, central venous (catheter) and arterial blood (indwelling catheter) samples were simultaneously collected, once per patient. Arterial measurements obtained with Precision PCx, and arterial, fingerstick and venous measurements obtained with Accu-chek Advantage II were compared to arterial central lab measurements. Agreement between point-of-care and laboratory measurements were evaluated with Bland-Altman, and multiple line...