five

Supplementary Material for: A new tool to assess patient-ventilator synchrony in preterm infants receiving non-invasive ventilation: a randomized crossover pilot study

收藏
DataCite Commons2025-05-01 更新2025-05-07 收录
下载链接:
https://karger.figshare.com/articles/dataset/Supplementary_Material_for_A_new_tool_to_assess_patient-ventilator_synchrony_in_preterm_infants_receiving_non-invasive_ventilation_a_randomized_crossover_pilot_study/28228892/1
下载链接
链接失效反馈
官方服务:
资源简介:
Introduction: Nasal synchronized intermittent positive pressure ventilation (nSIPPV) is an effective non-invasive ventilation technique, especially for preterm infants. Patient-ventilator synchrony is essential for providing effective respiratory support; however, no automated system is currently available for monitoring this parameter. A new tool for automatic assessment of patient-ventilator synchrony, the SyncNIV system, was developed and applied in this pilot study to evaluate differences between nSIPPV and non-synchronized nasal intermittent positive pressure ventilation (nIPPV) in preterm infants with respiratory distress. Methods: This study involved designing a custom algorithm for signal analysis. Data were collected through a polygraph that could simultaneously gather respiratory data from the patients and ventilator. Patient-ventilator synchrony was evaluated by applying the SyncNIV system in a randomized crossover study designed to compare nSIPPV and nIPPV. The primary outcome was the mean instant Synchrony Index (i-SI), defined as the portion of the inspiration effort sustained by ventilator inflation, expressed as a percentage. Results: Fourteen infants with a median (IQR) gestational age of 28.6 (25.6-30.3), were enrolled. We analyzed 43,304 ventilator inflations and 50,221 patient breaths. The i-SI was 54.69% (44.49-60.09) in nSIPPV and 39.54% (33.40-48.75) in nIPPV, p<0.05. Conclusion: The SyncNIV system confirmed better i-SI during nSIPPV than during nIPPV, demonstrating its effectiveness in assessing the differences between these two modes of non-invasive ventilation in preterm infants. The SyncNIV system could be a useful tool for optimizing the ventilation parameters and improving the effectiveness and comfort of respiratory support systems.
提供机构:
Karger Publishers
创建时间:
2025-01-17
二维码
社区交流群
二维码
科研交流群
商业服务