Characterizing x-ray transmission through filters used in high energy density physics diagnostics
收藏Mendeley Data2024-01-31 更新2024-06-27 收录
下载链接:
https://dataverse.harvard.edu/citation?persistentId=doi:10.7910/DVN/ROSRYW
下载链接
链接失效反馈官方服务:
资源简介:
We report on the design and implementation of a new system used to characterize the energy-dependent x-ray transmission curve, Θ(E), through filters used in high-energy density physics diagnostics. Using an Amptek X-123-CdTe x-ray spectrometer together with a partially depleted silicon surface barrier detector, both the energy spectrum and total emission of an x-ray source have been accurately measured. By coupling these detectors with a custom PROTO-XRD x-ray source with interchangeable cathodes, accurate characterizations of Θ(E) for filters of varying materials and thicknesses have been obtained. The validity of the technique has been confirmed by accurately reproducing areal densities for high-purity filters with known x-ray transmission properties. In this paper, the experimental setup is described and the results of absorption calibrations performed on a variety of different filters are presented.
创建时间:
2024-01-31