Carbon dioxide (CO2) densimeter calibration data
收藏DataONE2017-06-28 更新2025-04-20 收录
下载链接:
https://search.dataone.org/view/sha256:77a1a2305771c5a26d053b00b426b9d2c83cb2107cb54689c6b2d3b0cc64d975
下载链接
链接失效反馈官方服务:
资源简介:
Splitting of the carbon dioxide Raman Fermi diad was measured using three different Raman instruments and various lasers and dispersion gratings. Data were collected at Virginia Tech and at the U.S. Geological Survey in Reston, VA. The purpose of the study was to compare densimeters determined using different analytical methods. The results show that calibrations are instrument and analytical condition dependent, and one should develop a densimeter that is specific to the users instrument and analytical protocols to obtain carbon dioxide pressures based on Raman analysis of bubbles in fluid and melt inclusions.
创建时间:
2025-03-19