Pressure Correction in the Calibration of Heat Dissipation Sensors

收藏
NIAID Data Ecosystem2026-03-12 收录
下载链接:
http://datadryad.org/dataset/doi%253A10.5061%252Fdryad.mcvdncjwc
下载链接
链接失效反馈
官方服务:
资源简介:
Heat dissipation sensors (HDSs) infer soil matric potential based on the measured rate of temperature rise from a heat source. The process for calibrating these sensors may use a pressure plate apparatus (PPA) to remove water to a certain calibration point that is controlled by pressure. Conventionally, the pressure within the PPA is released before the temperature change calibration points are measured (DT0). This conventional HDS calibration approach can take several weeks. We propose that the equilibrated temperature change at pressure (DTp) can be corrected empirically to the value at zero pressure. Based on measurements of 14 HDSs between 0.02 and 0.1 MPa, DTp was on average 6.2% larger than DT0. After pressure correction, using an empirical equation, the difference between the corrected DTp and DT0 was only 0.28%. These results demonstrate that this empirical approach provides sufficient accuracy to achieve HDS calibration without performing a PPA pressure release.