IODP Expedition 367 Thermal conductivity
收藏NIAID Data Ecosystem2026-05-02 收录
下载链接:
https://zenodo.org/record/14179497
下载链接
链接失效反馈官方服务:
资源简介:
Thermal conductivity was measured using the heated needle method in either full-space needle configuration (soft/saturated sediments) or half-space needle configuration (harder materials) of a TeKa Berlin TK-04 thermal conductivity meter. Heat is applied to the sample and then thermal equilibrium is sought. The heating and equilibration curve is reduced to derive the thermal conductivity value. Raw data are stored if a user wishes to do off-line or postexpedition reduction.
创建时间: