遇见数据集

Dataset related to the article "Lesion Index: A Novel Guide in the Path of Successful Pulmonary Vein Isolation"

收藏
Zenodo2020-05-20 更新2026-04-07 收录
数据链接:
官方服务:

资源简介:

This record contains raw data related to the article "Lesion Index: A Novel Guide in the Path of Successful Pulmonary Vein Isolation". ABSTRACT <strong>Purpose: </strong>Previous studies indicate force time integral (FTI) as a radiofrequency (RF) lesion quality marker, while not considering power supply. Tacticath™ Quartz catheter provides Lesion index (LSI), a lesion quality marker derived by contact force (CF), power supply, and RF time combined. Our aim is to assess LSI and FTI correlation and a LSI-related cutoff of atrial fibrillation (AF) recurrences 12 months after pulmonary vein isolation (PVI). <strong>Methods: </strong>We retrospectively enrolled 37 patients who underwent RF ablation using Tacticath™ Quartz catheter. AF recurrence rate was evaluated 3, 6, and 12 months after PVI procedure. <strong>Results: </strong>AF recurrence was detected in 32% of patients. FTI mean value was significantly lower in left superior pulmonary vein (LSPV: 256 ± 86 gs vs 329 ± 117 gs, p = 0.05) and right inferior pulmonary vein (RIPV: 253 ± 128 gs vs 394 ± 123 gs p = 0.006) in patients with AF recurrences; no significant differences were found in right superior pulmonary vein (RSPV) and left inferior pulmonary vein (LIPV). LSI instead was significantly higher for all veins in patients without AF recurrences: LSPV (5.2 ± 0.7 vs 4.6 ± 0.8, p = 0.03), LIPV (5.0 ± 0.8 vs 4.5 ± 0.6, p = 0.04), RSPV (5.5 ± 0.6 vs 5.1 ± 0.6, p = 0.05), and RIPV (5.5 ± 0.7 vs 4.7 ± 0.8, p = 0.006). Receiver operator characteristic curve suggests 5.3 as LSI overall cutoff value predicting freedom from disease at 1-year follow-up. <strong>Conclusions: </strong>Our preliminary data suggest that a LSI mean value higher than 5.3 can be considered a good predictor of AF freedom at 1-year follow-up.

创建时间:
2020-05-20
二维码
社区交流群
二维码
科研交流群
商业服务