MicroED/3DED datasets of acetaminophen collected with a Ceta camera on Talos Arctica
收藏NIAID Data Ecosystem2026-05-02 收录
下载链接:
https://zenodo.org/record/14233389
下载链接
链接失效反馈官方服务:
资源简介:
Continuous rotation data sets were collected on a Talos Arctica (200 kev) equipped with a ceta camera. Each dataset was collected with an approximate flux of 0.17 e-/ Å2/s-1 which means a total exposure of 20.4 e-/Å2 for a 120-degree wedge of data. Delphi was used to rotate the crystal while TEM Imaging & Analysis (TIA) software was used to collect the data.
Datasets were processed in DIALS directly without any conversion. The structure was solved to 0.84 Å with 99% of overall completeness by merging two datasets (attached).
Experimental parameters are as follows:
Accelerating voltage: 200 Kev
Wavelength: 0.02508
Cameralength: 1012 mm
Exposure time: 1 s
Rotation speed: 1 degrees per second
Rotation: 60 to -60 degrees
Detector area: 4096 x 4096 (Bin 1)
Pixel size: 14 µm
Spot size: 8
C2 aperture: 50 µm
SA aperture: 40 µm
创建时间:
2024-11-28