ABSTRACT Objective: This study was aimed at evaluating the influence of cervical preflaring using LA-Axxes No. 1 bur (SybronEndo, Glendora, USA) or S1 and SX ProTaper files (Dentsply Maillefer, Ballaigues, Switzerland) on the accuracy of the Electronic Apex Locators (EALs): Root ZX mini (J. Morita Corporation, Tokyo, Japan), and Joypex 5 (Denjoy, Changsha, China). Methods: Thirty mandibular incisors were accessed, and the root canal length (RCL) was determined with a K-file #15, with the aid of a stereo microscope. Afterwards, the specimens were divided into 2 groups (n = 15) referring to the cervical preflaring with LA-Axxess or ProTaper files. The teeth were embedded in alginate and the RCL was determined by the EALs before and after preflaring. Data were classified into: accurate, if the difference in RCL measurement were ≤0.05 mm; and inaccurate, if the difference were >0.5 mm or beyond the RCL. Results: McNemar's test (α 0.05) after the cervical preflaring, regardless of the used instrument. Conclusion: The preflaring procedure increased the number of accurate measurements for both EALs, with statistical difference for Joypex 5 when the preflaring was performed with LA-Axxess. However, after the cervical preflaring, the EALs showed similar accuracy, regardless of the used instrument.