PEEK and CFR-PEEK knee wear dataset
收藏官方服务:
资源简介:
This study investigated the wear performance of PEEK and CFR-PEEK in a low conformity total knee replacement configuration. Custom-made flat inserts were tested against cobalt-chromium femoral bearings in a knee wear simulation for a period of three million cycles. Wear was assessed gravimetrically at intervals throughout the study. This dataset presents the input profile for the knee wear simulator, the experimental wear data and associated figures relating to the paper
本研究探究了低贴合型全膝关节置换构型下,聚醚醚酮(PEEK)与碳纤维增强聚醚醚酮(CFR-PEEK)的磨损性能。试验将定制扁平衬垫与钴铬股骨轴承组件置于膝关节磨损模拟装置中,开展时长为300万次循环的磨损测试。试验全程通过重量分析法定期评估磨损量。本数据集包含膝关节磨损模拟器的输入配置文件、试验磨损数据,以及与该论文相关的附属图表。
提供机构:
University of Leeds创建时间:
2017-04-06
搜集汇总
数据集介绍

背景与挑战
背景概述
该数据集聚焦于生物医学工程领域,研究了PEEK和CFR-PEEK材料在低配合度全膝关节置换配置中的磨损性能,通过300万次循环的膝关节磨损模拟实验,收集了模拟器输入参数、重量法磨损数据及相关图表。数据集提供了详细的实验数据和分析结果,旨在评估这些材料在膝关节植入物中的应用潜力,并附有相关学术论文链接,适用于材料科学和医疗设备研究。