Data for PSA 10.536 Elemental Hg generator performance evaluation report
收藏资源简介:
During the SI-Hg performance evaluation of elemental mercury gas generators on the market three generators were tested, e.g., PSA 10.536 elemental Hg generator, bell-jar and Tekran Model 3425. Key characteristics were determined e.g.; the stabilisation period, short-term drift, precision, i.e., reproducibility and repeatability of the concentration generated, linearity, bias, sensitivity to sample gas pressure, sensitivity to surrounding temperature and sensitivity to electrical voltage. All three generators could be tested according to the calibration protocol developed within the project. The results obtained with the different gas generator clearly shows the importance of a metrological calibration. All three candidate generators show a different bias for the setpoint compared to the calibrated output. The data obtained during the performance evaluation of the PSA 10.536 elemental Hg generator is published in this repository. The files of the following experiments can be found here: range1 m1 Calibration_PSA_range1_m1_20221130 multi_point_calibration_PSA_range1_m1 range1 m2 Calibration_PSA_range1_m2_20221209 multi_point_calibration_PSA_range1_m2 range1 m3 Calibration_PSA_range1_m3_20221214 multi_point_calibration_PSA_range1_m3 range1 m4 Calibration_PSA_range1_m4_20230915 multi_point_calibration_PSA_range1_m4 range1 m5 Calibration_PSA_range1_m5_20230919 multi_point_calibration_PSA_range1_m5 range2 m1 Calibration_PSA_range2 m1 20221006 multi_point_calibration_PSA_range2_m1 range2 m2 Calibration_PSA_range2 m2 20221011 multi_point_calibration_PSA_range2_m2 range2 m3 Calibration_PSA_range2 m3 20221012 multi_point_calibration_PSA_range2_m3 short-term drift m1 Calibration mercury gas generator range1 short term drift 20221012 deel 1 PSA_short_term_drift_M1 m2 Calibration mercury gas generator range1 short term drift 20221012 deel 2 PSA_short_term_drift_M2 m3 Calibration mercury gas generator range1 short term drift 20221012 deel 3 PSA_short_term_drift_M3 m4 Calibration mercury gas generator range1 short term drift 20221012 deel 4 PSA_short_term_drift_M4 stability Calibration mercury gas generator range2 stability 20220811 Calibration mercury gas generator stability 20221018
针对市售元素汞气体发生器开展国际单位制可溯源汞(SI-Hg)性能评估,本次测试共纳入3款商用发生器,分别为PSA 10.536型元素汞发生器、钟罩式发生器及Tekran Model 3425型发生器。本研究测定了三款设备的多项关键性能指标,包括稳定时长、短期漂移、浓度生成的精密度(即可重复性与复现性)、线性度、偏移量、样品气体压力灵敏度、环境温度灵敏度与电源电压灵敏度。所有三款发生器均可按照本项目内部制定的校准规程完成测试。不同气体发生器的测试结果清晰体现了计量校准的重要性:三款候选发生器的设定点与校准后输出值均存在不同程度的偏移。 本仓库公开了PSA 10.536型元素汞发生器性能评估过程中获取的实验数据,以下实验的相关文件均可在此获取: range1 m1:Calibration_PSA_range1_m1_20221130、multi_point_calibration_PSA_range1_m1 range1 m2:Calibration_PSA_range1_m2_20221209、multi_point_calibration_PSA_range1_m2 range1 m3:Calibration_PSA_range1_m3_20221214、multi_point_calibration_PSA_range1_m3 range1 m4:Calibration_PSA_range1_m4_20230915、multi_point_calibration_PSA_range1_m4 range1 m5:Calibration_PSA_range1_m5_20230919、multi_point_calibration_PSA_range1_m5 range2 m1:Calibration_PSA_range2 m1 20221006、multi_point_calibration_PSA_range2_m1 range2 m2:Calibration_PSA_range2 m2 20221011、multi_point_calibration_PSA_range2_m2 range2 m3:Calibration_PSA_range2 m3 20221012、multi_point_calibration_PSA_range2_m3 短期漂移(short-term drift): m1:Calibration mercury gas generator range1 short term drift 20221012 deel 1、PSA_short_term_drift_M1 m2:Calibration mercury gas generator range1 short term drift 20221012 deel 2、PSA_short_term_drift_M2 m3:Calibration mercury gas generator range1 short term drift 20221012 deel 3、PSA_short_term_drift_M3 m4:Calibration mercury gas generator range1 short term drift 20221012 deel 4、PSA_short_term_drift_M4 稳定性(stability):Calibration mercury gas generator range2 stability 20220811、Calibration mercury gas generator stability 20221018