遇见数据集

High-frequency measurements of chlorophyll fluorescence, characterizing F.I.Z. bias.

收藏
官方服务:

资源简介:

This dataset consists of high-frequency measurements of chlorophyll a fluorescence (Fchl), collected as part of manipulative experiments conducted in Lake George, NY. Experiments were set up to test for the potential of phototactic zooplankton to interfere with Fchl measurements. To test for any bias associated with fluorometer interference by zooplankton (FIZ), fluorometers were placed in the shallows of Lake George, and collected data under various treatment conditions. It was found that excitation light from fluorometers triggered a positive phototactic response during nighttime hours, biasing Fchl data by as much as 31x. Full results gleaned from this dataset can be found in: Moriarty, V.W., Lucius, M.A., Johnston, K.E., Borrelli, J.J., Mattes, B.M., Pezzuoli, A.R., Watson, C.D., Eichler, L.W. and Relyea, R.A. (2021), Fluorometer optical path interference via zooplankton phototaxis: Implications for high‐frequency data collection. Limnol Oceanogr Methods. https://doi.org/10.1002/lom3.10411