In vitro Volumetric Particle Velocimetry, Computational Fluid Dynamics (CFD), and in vivo 4D Flow MRI Hemodynamic Data in Two Patient-Specific Cerebral Aneurysms - CFD
收藏DataCite Commons2025-12-18 更新2024-07-13 收录
下载链接:
https://purr.purdue.edu/publications/3311/1
下载链接
链接失效反馈官方服务:
资源简介:
<p>This CFD dataset is part of a larger series that also includes MRI and STB datasets from the same study. The three sets can be downloaded separately, or in one bundle.</p>
<p>The data provided in this publication was used to evaluate parameters including wall shear stress (WSS), oscillatory shear index (OSI), and relative residence times (RRT) across multiple modalities and two patient-specific aneurysm geometries. Specifically, the assumptions and limitations of each modality coupled with the varying and often low spatial resolution and their subsequent effect on the hemodynamic metrics was explored. The results of this study demonstrated that non-dimensional parameters, such as OSI, were more robust to varying modalities and spatial resolutions and thus may provide more robust risk or growth and rupture assessments. These results demonstrate a need for further investigation and development of non-dimensional parameters and correlation of such parameters to risk of aneurysm.</p>
<p>Additionally, the pulsatile particle velocimetry results provided here can be used as robust validation test cases for computational fluid dynamics (CFD) solvers.</p>
<p>These data files were created using Matlab 2016.</p>
提供机构:
Purdue University Research Repository
创建时间:
2019-11-01