five

PFMCal: Photonic force microscopy calibration extended for its application in high-frequency microrheology

收藏
Mendeley Data2024-06-25 更新2024-06-26 收录
下载链接:
https://data.mendeley.com/datasets/s59f3gz729
下载链接
链接失效反馈
官方服务:
资源简介:
The present document is an update of the previously published MatLab code for the calibration of optical tweezers in the high-resolution detection of the Brownian motion of non-spherical probes. In this instance, an alternative version of the original code, based on the same physical theory, but focused on the automation of the calibration of measurements using spherical probes, is outlined. The new added code is useful for high-frequency microrheology studies, where the probe radius is known but the viscosity of the surrounding fluid maybe not. This extended calibration methodology is automatic, without the need of a user’s interface. A code for calibration by means of thermal noise analysis is also included; this is a method that can be applied when using viscoelastic fluids if the trap stiffness is previously estimated. The new code can be executed in MatLab and using GNU Octave. The previous version of this program (AEXH_v1_0) may be found at http://dx.doi.org/10.1016/j.cpc.2015.06.024

本文档为已公开发布的原MatLab代码的更新版本,该原代码用于高分辨率检测非球形探针布朗运动时的光镊(optical tweezers)校准。本更新版本推出了基于相同物理理论、但专注于球形探针测量校准自动化的原代码变体。新增代码适用于高频微流变学研究场景——此类场景中探针半径已知,但周围流体的黏度往往未知。该扩展校准方法可全自动运行,无需用户界面。此外还包含基于热噪声分析的校准代码:若此前已估算光阱刚度,该方法可用于粘弹性流体的相关实验。新版代码可在MatLab与GNU Octave环境中运行。该程序的旧版本(AEXH_v1_0)可通过链接http://dx.doi.org/10.1016/j.cpc.2015.06.024获取。
创建时间:
2024-01-23
5,000+
优质数据集
54 个
任务类型
进入经典数据集
二维码
社区交流群

面向社区/商业的数据集话题

二维码
科研交流群

面向高校/科研机构的开源数据集话题

数据驱动未来

携手共赢发展

商业合作