Dataset of PMU calibrator calibration.
收藏NIAID Data Ecosystem2026-05-01 收录
下载链接:
https://zenodo.org/record/6670031
下载链接
链接失效反馈官方服务:
资源简介:
Calibration data for calibration of Phasor Measurement Unit calibrator. Data includes GNU Octave script to process and display results.
The calibration method relies on the properties of a two channel digitizer with a time base common toboth ADCs. Because of the common time base, time errors of samples acquired by first channel are almost the same as the time errors of the samples acquired by second channel. The first channel is used to calibrate the time base and the second channel is used to sample the synchrophasor waveform.
Data are part of paper published in 25th IMEKO TC4 International Symposium, 23rd International Workshop on ADC and DAC Modelling and Testing, IMEKO TC-4 2020, Brescia, Italy / September 12-14, 2022.
Title: Simple method for calibration of PMU calibratorsAuthors: Martin Šíra and Stanislav MašláňContact: Czech Metrology Institute, Okružní 31, 638 00 Brno, msira@cmi.cz
Part of project Quantum traceability for AC power standards, QuantumPower, Project Number: 19RPT01. This project (19RPT01) has received funding from the EMPIR programme co-financed by the Participating States and from the European Union's Horizon 2020 research and innovation programme.
https://www.euramet.org/research-innovation/search-research-projects/details/project/quantum-traceability-for-ac-power-standards/?L=0&tx_eurametctcp_project%5Baction%5D=show&tx_eurametctcp_project%5Bcontroller%5D=Project&cHash=05c245c325fe8dcb6a358fe6e192e199