five

Supplementary Material for: Differential Molecular Modeling Predictions of Mid and Conventional Dialysate Flows

收藏
Figshare2019-01-30 更新2026-04-29 收录
下载链接:
https://figshare.com/articles/dataset/Supplementary_Material_for_Differential_Molecular_Modeling_Predictions_of_Mid_and_Conventional_Dialysate_Flows/7649444
下载链接
链接失效反馈
官方服务:
资源简介:
Background: High dialysate flow rates (QD) of 500–800 mL/min are used to maximize urea removal during conventional hemodialysis. There are few data describing hemodialysis with use of mid-rate QD (300 mL/min). Methods: We constructed uremic solute (urea, beta2-microglobulin and phosphate) kinetic models at varying volumes of distribution and blood flow rates to predict solute clearances at QD of 300 and 500 mL/min. Results: Across a range of volumes of distribution a QD of 300 mL/min generally yields a predicted urea spKt/V greater than 1.2 during typical treatment times with a small difference in urea spKt/V between a QD of 300 and 500 mL/min. A larger urea KoA dialyzer and 15 min of additional time narrows the urea spKt/V difference. No substantial differences were observed regarding the kinetics of beta2-microglobulin and phosphate for QD of 300 vs. 500 mL/min. Conclusion: A QD of 300 mL/min can achieve urea clearance targets. Hemodialysis systems using mid-rate QD can be expected to provide adequate hemodialysis, as currently defined.