遇见数据集

Supplemental material to the article "Sub-nm precise measurement of adhesive gaps by imaging interferometry"

收藏
DataCite Commons2024-11-10 更新2025-04-16 收录
官方服务:

资源简介:

This dataset contains experimental data from length measurements of adhesive gaps in substrates with and without adhesives at different pressure, temperature, air moisture and time. The measurements are performed using PTB's Precision Interferometer on a standard material (ULE®) together with a common adhesive (3MTM Scotch-WeldTM DP-460EG), gap thickness approx. 100 µm. Measurements on the sample without adhesive reveal that the length changes without adhesive are within approx. 0.1 nm under comparable conditions. On the adhesive sample, on the other hand, drying in a vacuum and water vapor treatment lead to shrinkage and swelling. These long-term process processes can be described very well by simple but empirical fitting functions. The measurement uncertainty budget results in typical values of approx. 0.3 nm for changes of the adhesive gap thickness. This is orders of magnitude more accurate than was previously possible using traditional methods on adhesive samples.

创建时间:
2024-08-20
二维码
社区交流群
二维码
科研交流群
商业服务